15 Maintenance
98
7
Lift off the back of the instrument.
Assembly
Perform in reverse order to assemble. Please note:
- Lay hoses and lines in the guides intended for this purpose.
- Make sure that hoses and lines do not get jammed.
15.7 Replace sensors
A slot bridge (0192 1552) must be inserted in slots which are not
equipped with a sensor. Used sensors must be disposed of as
hazardous waste!
Available slots:
1
CO sensor or COlow sensor
O2 sensor
2
NO sensor or NOlow sensor
When retrofitting a sensor, the associated measurement parameter/unit
must be enabled in the reading display.