14 Performing the measurement
76
Edit values
All values that can be modified have a dotted underlining.
The function is only available if the chosen fuel is an oil.
This function is used to calculate the capacity of the burner from the set oil
pressure and the oil flow rate of the oil nozzle.
Call up function: | Oil flow rate
Set oil flow rate of the oil nozzle and oil pressure.
The calculated oil burner capacity is displayed (in KW).
All values that can be modified have a dotted underlining.
• Cigarette smoke influences the measurement by more than 50 ppm.
The breath of a smoker influences the measurement by about 5
ppm.
• When using an ambient CO probe, note that:
The direction of flow of the gas has an effect on the accuracy of
measurement. Frontal flow onto the sensor leads to higher
readings. The best measurement results are achieved when the
probe is moved gently backwards and forwards.
• When using the ambient CO probe and the flue gas probe, note
that:
The probe must be in the fresh air (CO-free) during the zeroing
phase.
For this type of measurement, the Mean value calculation option is
also available. See Mean value calculation section.
An ambient CO probe (0632 1272) must be connected.
Call up function: | CO Ambient