15 Maintenance
99
15.8 Replace O2 sensor
CAUTION
Acid in the sensors.
May cause chemical burns.
- Do not open the sensors.
- Always wear gloves when changing a sensor.
Eye contact: Rinse the affected eye thoroughly under running water for 10
minutes, keeping the eyelids wide open and protecting the unaffected eye.
Remove contact lenses wherever possible.
Measuring instrument is open, see Section Open measuring
instrument.
Unlock retaining bracket and open
out.
Remove faulty sensor from the slot.
Insert new sensor in the slot.
Make sure that the socket on the
sensor circuit board is correctly
connected to the contact plug.
Close the retaining bracket with an audible "click".