11 Performing the measurement
31
o Coldspot: the lowest temperature measuring point is
marked with blue crosshairs and the value is displayed.
o Show all / Hide all: Show or hide the centre point, hotspot
and coldspot.
• New Min/Max on area:
o Min/Max on area: Area in the centre of the image is
shown. The minimum, maximum, and average values for
the area are displayed.
o Hotspot: the highest temperature measuring point within
the range selection is marked with blue or red crosshairs
and the value is displayed.
o Coldspot: the lowest temperature measuring point within
the range selection is marked with blue or red crosshairs
and the value is displayed.
o Show all / Hide all: Show or hide the selected area.
• Measurement range: choose between two measuring ranges
or temperature ranges.
• Differential temperature: identifies the difference between two
temperatures.
o Difference between two measuring points
o Difference between measuring point and input value
o Difference between measuring point and value of an
external probe
o Difference between measuring point and the reflected
temperature (RTC)
• External values: in various measuring modes, values can be
determined either manually or with a Bluetooth
®
measuring
instrument.
• iFOV: the IFOV warner shows what can be measured
accurately from a specific distance.
• Isotherm: limit values can be set. All readings within the set
limits are displayed uniformly in one colour.
• Alarm: Temperatures below or above a limit value are also
visualized.
• Zoom: enlarges the image section (2x and 4x).
• SiteRecognition: testo SiteRecognition recognizes markers,
the recorded images are assigned to the corresponding
measuring location by the IRSoft PC software and stored in the
database.
Select the required function (joystick or touchscreen).