12 Maintenance
84
Click on Configure thermal imager.
The Camera settings window is opened.
Select Instrument settings -> Firmware Update -> OK.
FW_T883_Vx.xx.bin is displayed.
Firmware update is performed.
Thermal imager is automatically restarted. Update complete.
In certain circumstances, the camera reports: Firmware update
finished. Please restart the device. A restart is not performed.
Action:
- After 10 seconds, switch off the camera; switch back on again after 3
seconds.
The current firmware version is displayed.
12.4.2 Performing an update with the camera
12.4.2.1 Preparing the camera
The rechargeable battery is fully charged or the camera is connected to
the mains adapter.
Open the cover of the interface.
Connect the computer to the camera with the USB cable.
The camera is displayed as a removable medium in Windows Explorer.
12.4.2.2 Performing an update
Copy the FW_T883_Vx.xx.bin file to the removable medium using
drag & drop.
Eject the removable medium.
Disconnecting the USB connection.