1. Remove the kick plate at the bottom of the cabinet by removing the
two Phillips screws indicated by the black arrows in the illustration.
Disconnect the cable inside the cabinet.
2. Open the outer door and remove it by lifting it off its hinge pins. Place
the door on a padded surface to prevent scratches (Figure 5-2).
Note The hinge pins are not connected to the brackets and may fall out
when the door is lifted off.
s
3. Refer to the illustration of the outer door in Figure 5-2. With a
Phillips screwdriver, remove and save the four nylon screws from the
door hinge mounting holes. The black arrows identify the screw
locations.
4. Remove the door hinges, rotate them 180° and secure them on the left
side of the door. Install the nylon screws into the newly exposed holes.
5. Remove the two plastic clips securing the door heater cable. These clips
will be rotated 180° when installed on the left lower edge of the door
frame.
5-4 Model 310 Series DH Incubator Thermo Scientific
Section 5
Routine Maintenance
Figure 5-2. Remove Two Screws
Reversing the Door
Swing (continued)