EasyManua.ls Logo

Thermo Scientific 310 - Page 61

Thermo Scientific 310
86 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Figure 6-4. Connector Terminals
4. Each system monitored (Temp, CO
2, RH) requires two conductors
(refer to Figure 6-4). Feed the wires through the auxiliary wire
through-port on the back panel. This port is a circular fitting adjacent
to the remote alarm and RS485 connectors as illustrated in Figure 6-3.
The wire routing is shown by the black arrows in Figure 6-5.
5. Strip the ends of each conductor and wire it to the appropriate
terminals of connectors J2 and/or J3 on the analog board. Refer to
Figure 6-4.
6. When wiring is completed, slide in the electronics drawer or the
replace the sheet metal cabinet top.
7. Replace the screws removed earlier and return the unit to service.
6-4 Model 310 Series DH Incubator Thermo Scientific
Section 6
Factory Options
Figure 6-5. Wire Routing
Connect Analog
Output Boards (cont.)

Table of Contents

Related product manuals