6
Maintenance
Adding Calibration Compound
102 TSQ Quantum XLS Series User Guide Thermo Scientific
Adding Calibration Compound
The calibration compound is a liquid whose mass spectrum of ions is used to tune and
calibrate the TSQ Quantum XLS Series mass spectrometers. The mass spectrometers use
FC-43 as their calibration compound. Although you cannot visually determine when to add a
calibration compound, Xcalibur Automatic Tune produces the error message: “the intensity of
calibration gas ions is too low.” You must add calibration compound; however, many factors
other than a decrease in the calibration compound can cause a shortage of calibration gas ions.
As a rule, add calibration compound once a year.
The calibration compound resides in a cylindrical vial (see Figure 56).
Figure 56. Calibration gas flow modules, showing the calibration compound vial
Y To add calibration compound
1. In EZ Tune, place the instrument in standby.
2. Remove the cover to the calibration compound vial compartment. See Figure 57.
Tools needed
Calibration compound (P/N 50010-02500)
Syringe (P/N 36502019)
Frequency
Yearly or as needed
Calibration compound vial

Table of Contents

Other manuals for Thermo Scientific TSQ Quantum XLS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific TSQ Quantum XLS and is the answer not in the manual?

Thermo Scientific TSQ Quantum XLS Specifications

General IconGeneral
BrandThermo Scientific
ModelTSQ Quantum XLS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals