EasyManuals Logo

Thermo Scientific TSQ Quantum XLS User Manual

Thermo Scientific TSQ Quantum XLS
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #139 background imageLoading...
Page #139 background image
7
Diagnostics and Troubleshooting
Troubleshooting
Thermo Scientific TSQ Quantum XLS Series User Guide 123
Issue: Vacuum is faulty. Possible Solutions
A gross leak is present. Check for leaks.
The pressure in the analyzer region of the vacuum manifold, as measured by the ion
gauge, must be below the pressures listed in Table 3 on page 55.
Linearity Issues
Linearity issues occur when a plot of intensity versus concentration of a known compound is
not linear. Poor instrument operating conditions can cause linearity problems. In addition,
certain compounds do not give a desired linear response due to chromatographic activity.
A well-maintained instrument provides good linear response over a wide range of
concentrations for most compounds. Like all instruments, however, the TSQ Quantum XLS
Series mass spectrometers have a saturation point.
Perform routine injector and column maintenance to minimize linearity problems. Usually, a
hardware fault that affects linearity shows different issues than those that might be attributed
to linearity.
Issue: Calibration plots
not sufficiently linear.
Possible causes/solutions
High-end standards are too concentrated for the MS. Use the split injection technique to
decrease the amount of sample, or lower emission current to reduce MS sensitivity.
The ion volume and lenses are dirty. Clean ion volume and lenses as described in
“Cleaning Ion Source Componentson page 85.
The electron multiplier setting is incorrect. Run the tune and calibration procedure as
described in “Running Auto Tune and Calibrationon page 64.
The injection port liner or capillary column is dirty. Change the injection port liner and
trim the capillary column.
The capillary column stationary phase is too thin for high concentration samples. Use a
higher capacity capillary column with a thicker stationary phase, or use a split injection
technique.
The capillary column is bad. Replace the capillary column. See “Removing and Installing
a GC Capillary Columnon page 107.

Table of Contents

Other manuals for Thermo Scientific TSQ Quantum XLS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific TSQ Quantum XLS and is the answer not in the manual?

Thermo Scientific TSQ Quantum XLS Specifications

General IconGeneral
BrandThermo Scientific
ModelTSQ Quantum XLS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals