EasyManuals Logo

Thermo Scientific TSQ Quantum XLS User Manual

Thermo Scientific TSQ Quantum XLS
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #138 background imageLoading...
Page #138 background image
7
Diagnostics and Troubleshooting
Troubleshooting
122 TSQ Quantum XLS Series User Guide Thermo Scientific
The most reliable way to find vacuum leaks is to spray a gas around the vacuum manifold and
look for the characteristic ions in full-scan EI mode. Argon produces m/z 40. Or, you can use
compressed electronic dusting spray containing an HFC. For example, Falcon™ Dust-Off™
and MicroCare™ Micro-Blast™ contain tetrafluoroethane, which produces ions at m/z 69
and 83.
Issue: Forepump will
not turn on.
Possible causes/solutions
The forepump is off. Check the forepump switch.
The vacuum service switch is in the Off position. Place the vacuum service switch in the
operating position.
The forepump power cable from the mass spectrometer is not connected. Connect the
power cable.
The forepump is faulty. Replace the forepump.
Issue: Forepump
powers on, but will not
pump down.
Possible causes/solutions
The oil level in the forepump is insufficient. Check oil level; add oil if necessary.
The foreline is leaking. Check the clamps and connectors. Replace the foreline hose if a
hole is found.
The vacuum manifold is leaking.
The forepump is faulty. Replace the forepump.
Issue: Turbo pump
shuts off during
operation.
Possible cause/solution
The foreline pressure is too high. Check for leaks in the foreline.
Issue: Unexpected full
ventilation occurs.
Possible causes/solutions
A gross leak is present. Check for leaks.
The system was vented through the inlet valve. Close and plug the inlet valve.
The GC column broke at the transfer line. Replace the GC column. See “Removing and
Installing a GC Capillary Columnon page 107.
The foreline was cut. Replace the foreline hose.
Note The mass spectrometer performs a leak check during the automatic tuning and
calibration procedure and during the maintenance tune procedure. The mass
spectrometer returns a pass or fail report.

Table of Contents

Other manuals for Thermo Scientific TSQ Quantum XLS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific TSQ Quantum XLS and is the answer not in the manual?

Thermo Scientific TSQ Quantum XLS Specifications

General IconGeneral
BrandThermo Scientific
ModelTSQ Quantum XLS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals