4
Tuning and Calibrating
Saving the Tune and Calibration Report
Thermo Scientific TSQ Quantum XLS Series User Guide 67
7. Save the calibration file as follows:
a. Click Save Calib. As to open the Save Calibration File dialog box.
b. In the File Name box, enter a name for your calibration file.
c. Click Save to save the calibration file. The Save As dialog box appears.
8. Save the tune method file as follows:
a. In the File Name box, enter a name for your tune method file.
b. Click Save to save the tune method file.
The mass spectrometer is now tuned and calibrated in the EI positive ion mode.
Saving the Tune and Calibration Report
You can save the information that appears in the Status box and in the parameter optimization
plots as a PDF file. Examples of status information and parameter optimization plots from the
tune and calibration report follow in Figure 37, and Figure 38 on page 69.