7
Diagnostics and Troubleshooting
Troubleshooting
116 TSQ Quantum XLS Series User Guide Thermo Scientific
Issue: Acquisition does
not begin as expected.
Possible causes/solutions
The start mode is not properly configured. Verify that the instrument is configured
properly. Refer to the Xcalibur Help for more information.
The mass spectrometer Instrument Setup file has incorrect settings. Check the start time
in the Instrument Setup file.
The forepressure is too high due to the solvent peak. Reduce injection volume or extend
the filament/multiplier delay time until after solvent peak.
The remote start cable to the GC is disconnected. Connect the remote start cable.
The GC did not start. Verify GC instrument method and configuration. Verify the
connection between the autosampler and GC.
The autosampler did not start. Verify autosampler instrument method and configuration.
Verify that the sample is present.
The heated zone stipend is not attained. See “Heated Zone Issueson page 120.
Issue: Acquisition
terminates
unexpectedly.
Possible causes/solutions
The end run time in the TSQ Quantum XLS Series instrument method is incorrect.
Check the instrument method for the GC and the MS.
The data system is out of disk space. Check disk space; back up and remove files.
Issue: Unable to
initialize the gas
chromatograph.
Possible causes/solutions
The GC is not turned on. Turn on the GC.
The GC is not configured properly in the Xcalibur data system. Check Instrument
Configuration.
The cable between COM1 and GC is disconnected. Verify connection.
COM1 is not configured properly. Verify COM port configuration.
Issue: Unable to
download methods to
the gas chromatograph.
Possible causes/solutions
There is a discrepancy between the instrument method and configuration. Verify
consistency between method and instrument configuration.
COM1 is not configured properly. Verify COM port configuration.

Table of Contents

Other manuals for Thermo Scientific TSQ Quantum XLS

Questions and Answers: