Operation
Service Menu
Thermo Electron Corporation Model 42i Low Source Instruction Manual 3-87
Input Board Calibration The Input Board Calibration menu is used to initiate a calibration of the
input A/D stages. The input board calibration menu is visible only when
the instrument is in service mode. For more information on the service
mode, see “Service Mode” on page.
In the Main Menu, choose Service > Input Board Calibration.
Manual Input Calibration The Manual Input Calibration screen is used to do a manual calibration of
the input board A/D stages.
Note The measurement system and the PMT are both shut off inside this
screen.
In the Main Menu, choose Service > Input Board Calibration > Manual
Input Cal.
Press to leave warning screen.
Make a note of the frequency at gain of 1.
•Use
and to change the gain between 10 and 100.
Use and to increment or decrement the D/A counts so
the frequency at gain 100 is equal to the frequency at gain 1.
WARNING This adjustment should only be performed by an instrument
service technician.
INPUT BOARD CALIBRATION:
>MANUAL INPUT CAL
AUTOMATIC INPUT CAL
INPUT FREQUENCY DISP
RANGE AVG DIAGS ALARM
WARNING This adjustment should only be performed by an instrument
service technician.