Servicing
Ozonator Assembly Replacement
7-24 Model 42i Low Source Instruction Manual Thermo Electron Corporation
6. Re-install the measurement bench. Refer to “Removing the
Measurement Bench and Lowering the Partition Panel” in this chapter.
Figure 7-13. Replacing the Solenoid Valve
Ozonator Assembly
Replacement
Use the following procedure to replace the ozonator assembly (Figure 7-14).
Equipment Required:
Ozonator assembly
Wrench, 5/8-inch
Philips screwdriver
1. Refer to “Removing the Measurement Bench and Lowering the
Partition Panel” in this chapter to lower the partition panel, then
proceed to the next step below.
2. Carefully disconnect the plumbing at the glass inlet and outlet of the
ozonator.
Solenoid Valve
Mounting Clip
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.