C-Link Protocol Commands
Communications Configuration
B-38 Model 42i Low Source Instruction Manual Thermo Electron Corporation
where xxxx = 4 hexadecimal digits that represent the sum of all the
characters (bytes) in the message
host name
This command reports the host name string.
Send: host name
Receive: host name ISERIES
set host name string
This command sets the host name string, where string is 1-13 alphanumeric
characters.
Send: set host name analyzer01
Receive: set host name analyzer01 ok
instr name
This command reports the instrument name.
Send: instr name
Receive: instr name
NO-NO2-NOx Analyzer
NO-NO2-NOx Analyzer
instrument id
This command reports the instrument id.
Send: instrument id
Receive: instrument id 42
set instrument id value
This command sets the instrument id to value, where value is a decimal
number between 0 and 127 inclusive.
Note sending this command via RS-232 or RS-485 will require the host to
use the new id for subsequent commands.
Send: set instrument id 50
Receive: set instrument id 50 ok
Table B-9. Reply Termination Formats
Format Reply Termination
00 <CR>
01 <NL> sum xxxx <CR>

Table of Contents