C-Link Protocol Commands
Keys/Display
B-26 Model 42i Low Source Instruction Manual Thermo Electron Corporation
set no bkg value
set nox bkg value
These commands are used to set NO and NO
x
backgrounds to user-defined
values to value, where value is a floating-point representation of the
background in current selected units. The example below sets the NO
background to 5.5 ppb.
Send: set no bkg 5.5
Receive: set no bkg 5.5 ok
pres cal
This command reports the pressure recorded at the time of calibration. The
example below shows that the pressure at calibration is 185.5 mmHg.
Send: pres cal
Receive: pres cal 185.5 mmHg
set pres cal
This command automatically sets the current pressure as the calibration
pressure. The example below successfully sets the calibration pressure to
120.5 mmHg.
Send: set pres cal 120.5
Receive: set pres cal 120.5 ok
set cal pres
This command automatically sets the current pressure as the calibration
pressure. The example below successfully sets the calibration pressure.
Send: set cal pres
Receive: set cal pres ok
Keys/Display push button
button = | do | down | en | enter | he | help | le | left | me | menu | ri | right |
ru | run | up | 1 | 2 | 3 | 4 |
These commands simulates pressing the front panel pushbuttons. The
numbers represent the front-panel soft keys, from left to right.
Send: push enter
Receive: push enter ok