Thermo 42i - Pump Rebuilding

Thermo 42i
314 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Preventive Maintenance
Pump Rebuilding
Thermo Electron Corporation Model 42i Low Source Instruction Manual 5-5
3. Re-install the filters and fan guards.
Pump Rebuilding Use the following procedure to rebuild the pump (Figure 5-2). To replace
the pump. See “Pump Replacement” in the “Servicing” chapter.
Equipment Required:
Pump Repair Kit
Allen Wrench, 3 mm
Wrench, 9/16-inch
Needlenose Pliers
1. Turn instrument off, unplug the power cord, and remove the cover.
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component. For more information
about appropriate safety precautions, see the “Servicing” chapter.