Operation
Service Menu
Thermo Electron Corporation Model 42i Low Source Instruction Manual 3-93
Analog Input Calibrate Full-Scale The Analog Input Calibration Full-Scale screen allows the user to calibrate
the full-scale state of the selected analog input.
In the Main Menu, choose Service > Analog Input Calibration > Select
Channel > Calibrate Full Scale. (Hook up a voltage source of 10 V to
the analog input channel.)
Use and to increment or decrement the numeric value.
Press to save the value.
ANALOG INPUT CAL: ZERO
DISCONNECT SELECTED INPUT!
SELECTED INPUT: INPUT 1
CURRENTLY: 6.24 V
CALIBRATE INPUT TO ZERO
RANGE AVG DIAGS ALARM
ANALOG INPUT CAL: SPAN
PROVIDE VOLTAGE TO INPUT!
SELECTED INPUT: INPUT 1
CURRENTLY: 6.24 V
SET TO: 10.00 V
CALIBRATE TO VALUE
RANGE AVG DIAGS ALARM