C-Link Protocol Commands
Hardware Configuration
Thermo Electron Corporation Model 42i Low Source Instruction Manual B-33
conv set temp
This command reports the temperature that the NO
2
converter is set to.
The example below reports that the converter temperature is set to 625 °C.
Send: conv set temp
Receive: conv set temp 625 deg C
set conv set temp value
This command sets the temperature that the NO
2
converter is set to, where
value is an integer representing degrees C. The example below sets the
converter temperature to 625 °C.
Send: set conv set temp
Receive: set conv set temp 625 deg C ok
date
This command reports the current date. The example below reports the
date as December 1, 2004.
Send: date
Receive: date 12-01-04
set date mm-dd-yy
mm = month
dd = day
yy = year
This command sets the date of the analyzers internal clock. The example
below sets the date to March 19, 2005.
Send: set date 03-19-05
Receive: set date 03-19-05 ok
set default params
This command sets all the parameters to their default values. This does not
affect the factory-calibrated parameters.
Send: set default params
Receive: set default params ok
ozonator
This command reports the ozonator on or off. The example below reports
that the ozonator is on.
Send: ozonator
Receive: ozonator on