Operation
Calibration Menu
Thermo Electron Corporation Model 42i Low Source Instruction Manual 3-27
Next Time The Next Time screen is used to view and set the next zero/span check date
and time. Once the initial zero/span check is performed, the date and time
of the next zero/span check is calculated and displayed.
• In the Main Menu, choose Calibration > Zero/Span Check > Next
Date/Time.
• Use , , and to move and change the value of
the date and time.
•Press to accept a change.
Period Hours The Zero/Span Period Hours screen defines the period or interval between
zero/span checks. Periods between 0 and 999 hours are acceptable. To turn
the zero/span check off, set the period to 0.
• In the Main Menu, choose Calibration > Zero/Span Check > Period
Hours.
• Use and to increment and decrement the numeric value.
•Press to save the period.
NEXT DATE AND TIME: NEXT DATE AND TIME:
01 JAN 2005 12:00 02 JAN 2005 12:00 ?
PRESS
TO EDIT
SETTING: DAYS
¨ SET MONTHS
©ª CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
RANGE AVG DIAGS ALARM
ZERO/SPAN PERIOD:
CURRENTLY: 024 HRS
SET TO: 025 HRS ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM