Operation
Diagnostics Menu
Thermo Electron Corporation Model 42i Low Source Instruction Manual 3-67
Interface Board Voltages The Interface Board screen (read only) is used to display the current voltage
readings on the interface board.
• In the Main Menu, choose Diagnostics > Voltages > Interface Board
Voltages.
I/O Board Voltages The I/O Board screen (read only) is used to display the current voltage
readings on the I/O expansion board. This screen is displayed if the I/O
expansion board option is installed.
• In the Main Menu, choose Diagnostics > Voltages > I/O Board
Voltages.
MOTHERBOARD VOLTAGES:
3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
15.0 SUPPLY 15.0 V
24.0 SUPPLY 24.0 V
-3.3 SUPPLY -3.3 V
RANGE AVG DIAGS ALARM
INTERFACE BOARD VOLTAGES:
PMT SUPPLY 785.5 V
3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
15.0 SUPPLY 15.0 V
P15.0 SUPPLY 15.0 V
24.0 SUPPLY 24.0 V
-15.0 SUPPLY -15.0 V
RANGE AVG DIAGS ALARM
I/O BOARD VOLTAGES:
3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
15 0 SUPPLY 15.0 V
24.0 SUPPLY 24.0 V
-3.0 SUPPLY -3.0 V
RANGE AVG DIAGS ALARM