Detecting hot pixel automatically
opens the "Badpixel detection" dialog. Bad pixels are detected and can be added to
the "user" class list.
"Threshold": Any pixel with an intensity exceeding that threshold will be recognized as a bad (hot)
pixel, where the threshold is is the minimum intensity difference to the reference intensity. The
reference is the averaged over the most recent 10 frames total pixel intensities. Threshold can be
set between 5 and 255.
The "Clear old badpixel list" option removes the old user list before the search.
Note
The number of bad pixels, stored to the EEPROM, is limited. If the number of detected bad pixels
exceeds the memory capacity, please increase the bad pixel detection threshold and repeat the
detection procedure.