58
OBJECTIVE OPERATIONS
9 Using the control lever, move the instrument body and put the alignment bright spot into the ( )
scale.
10 When the positional relation between the instrument body and the patient's eye is proper for pho-
tography after putting the alignment bright spot into the ( ) scale, the fundus photography is auto-
matically done by the auto shoot function. Ask the patient not to blink and not to move his/her
eye during photography.
NOTE
By this operation, focusing and positioning are performed. However, photo-
graphing is not performed.
By this operation, the alignment bright sport is moved to the left.
When the auto focus function is ON, even if you pull the base unit toward the
patient in stereoscopic photography, the focusing knob is not reset to the initial
position.
When the auto focus function is OFF, turn the focusing knob to change the
split lines into one line.
When the auto shoot function is OFF or when the diopter compensation lens
is set to others except "0", put the alignment bright spot into the ( ) scale and
then select the .
NOTE
If the patient blinks right before the fundus photography, the automatic blink
detection function stops the photography. Start the procedure again from
alignment and take a picture.
When the auto shoot function is OFF or when the diopter compensation lens
is set to others except "0", the instrument does not take a picture automati-
cally. Press the to take a color picture of fundus.
STEREOSCOPIC PHOTOGRAPHY SELECTOR BUTTON
Stereoscopic
photography
selector button
PHOTOGRAPHY BUTTON

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Topcon DRI OCT-1 Triton and is the answer not in the manual?

Topcon DRI OCT-1 Triton Specifications

General IconGeneral
BrandTopcon
ModelDRI OCT-1 Triton
CategoryMedical Equipment
LanguageEnglish

Related product manuals