95
BEFORE REQUESTING SERVICE
FPGA initialization error.
Please turn the power switch OFF.
[210108]
Initialization of the built-in FPGA has failed.
Turn off the power and contact the service engineer.
Lamp house cover is off.
Please turn the power switch OFF and close the
cover.
[209109]
It is probable that the lamp house cover comes off or is not
attached properly.
Turn off the power and contact the service engineer.
FAN error.
Please turn the power switch OFF.
[209110/211110]
The FAN rotary speed is not normal. The temperature in the
instrument may be higher.
Turn off the power and contact the service engineer.
Failed to detect the center position of Ystage.
Please reboot.
[311202]
An error has occurred while the Y stage is initially operating.
Restart the instrument. Contact the service engineer for
inspection.
EEPROM data error.
Please reboot.
[309204/310204]
The EEPROM data in the built-in PCB is not normal.
Restart the instrument. Contact the service engineer for
inspection.
TRC focus motor initialization error.
Please reboot.
[309205]
An error has occurred while the TRC focus motor is initially
operating.
Restart the instrument. Contact the service engineer for
inspection.
Split unit motor initialization error.
Please reboot.
[309206]
An error has occurred while the split unit is initially operat-
ing.
Restart the instrument. Contact the service engineer for
inspection.
Diopter scale lens initialization error.
Please reboot.
[309207]
An error has occurred while the diopter compensation lens
motor is initially operating.
Restart the instrument. Contact the service engineer for
inspection.
Small pupil diaphragm motor initialization error.
Please reboot.
[309208]
An error has occurred while the small pupil diaphragm motor
is initially operating.
Restart the instrument. Contact the service engineer for
inspection.
Split initialization error.
Please reboot.
[309209]
An error has occurred while the split motor is initially operat-
ing.
Restart the instrument. Contact the service engineer for
inspection.
OCT focus motor initialization error.
Please reboot.
[310210]
An error has occurred while the OCT focus motor is initially
operating.
Restart the instrument. Contact the service engineer for
inspection.
Optical path correction motor initialization error.
Please reboot.
[310211]
An error has occurred while the light path length compensa-
tion motor is initially operating.
Restart the instrument. Contact the service engineer for
inspection.
Polarizer motor initialization error.
Please reboot.
[310212]
An error has occurred while the polarization controller motor
is initially operating.
Restart the instrument. Contact the service engineer for
inspection.
Optical path switching motor initialization error.
Please reboot.
[312214]
An error has occurred while the light path switching motor is
initially operating.
Restart the instrument. Contact the service engineer for
inspection.
Error message Contents

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Topcon DRI OCT-1 Triton and is the answer not in the manual?

Topcon DRI OCT-1 Triton Specifications

General IconGeneral
BrandTopcon
ModelDRI OCT-1 Triton
CategoryMedical Equipment
LanguageEnglish

Related product manuals