TROUBLESHOOTING
TROUBLESHOOTING
MESSAGE LIST
"OVER-SPH" Spherical power exceeds +22D or -25D.
"OVER-CYL" Cylindrical power exceeds ±10D.
"OVER-R" Corneal curvature exceeds 5.00-10.00mm.
"NO TARGET" There is no target or the eye image is too dark.
"AGAIN" There is more than ±5D difference from the previous measure-
ment value.
"NO CENTER" Center of eye can not be found.
"ALIGN ERR" Alignment significantly failed during the measurement.
"ERROR" The patient's eye blinks or moves during measurement.
If this message appears while with measuring model eye, the
instrument may have a problems. Contact your service engineer.
"LAN hostname Error" Failed in host name resolution of the destination (to be connected
with the share folder). Confirm the inputted host name or DNS
server address.
"LAN mount Error" Failed in connection with the share folder.
Confirm the address, folder name, user name and password of the
destination (to be connected with the share folder).
"LAN create Error" Failed in file creation.
Confirm that write permission to the share folder is set correctly.
"LAN write Error" Failed in writing to the file.
Check the free space capacity at the save location.
"RS232C FAIL" Failed in RS232C data transmission.
"Please check the
DATE/TIME"
The battery for the buit-in clock become run down. Before using,
confirm the time and date on the SETUP menu.
If the message comes up frequently, call your service engineer.
"Previous measurements
are left. Please press the
Clear button."
Displayed when the output of all output-set data fails.