48
SETTING FUNCTIONS ON SETUP SCREEN
LIST OF SETUP ITEMS
Setup items are categorized into 6 large indexes.
"Initial" .....................items related to the initial status after power on
"Print" ......................items related to output from the internal printer
"Comm" ...................items related to data output with the external device
"LAN".......................items related to output using the LAN
"Operator ID"...........items related to Operator ID.
"Special"..................items related to maintenance (for service engineer only)
INITIAL (INITIAL SETTING)
Initial contains settings related to the initial status after power on, clearing all measure-
ment values, etc.
Descriptions Options Details Initial value
Buzzer
OFF Buzzer does not sound.
ON
ON Buzzer sounds.
Start mode
MANUAL Default measurement mode is MANUAL.
AUTO
AUTO Default measurement mode is AUTO.
Auto print
OFF Not printed automatically.
ON
ON
After AUTO measurement,
results are printed out automatically.
Printer
OFF Internal printer is disabled.
ON
ON Internal printer is active.
Patient No. reset
OFF Patient No. is not reset upon power on.
OFF
ON Patient No. is reset upon power on.
Show patient ID
OFF Patient ID is not displayed.
ON
ON Patient ID is displayed.
Required
patient ID
OFF Patient ID is not displayed.
OFF
ON Patient ID is displayed.
Device ID
number
1-99
Set by ten-key display.
Sets the Device ID number. 1
Show Device ID
number
OFF Device ID is not required.
OFF
ON Device ID is required.
Stand by mode
RIGHT Waiting at the initial position for right eye measurement.
RIGHTLEFT Waiting at the initial position for left eye measurement.
LAST Waiting at the last position of the measured eye.
Auto power save
OFF Power save function is not used.
10min
1min Power save status in 1min after last operation.
5min Power save status in 5min after last operation.
10min Power save status in 10min after last operation.
20min Power save status in 20min after last operation.
30min Power save status in 30min after last operation.
60min Power save status in 60min after last operation.
Cont. Cycle
1-10
Set by ten-key display
The number of continuous measurements. 3
Add Measure
0-99
Set by ten-key display.
When the measurement is error,
set the number of times of remeasurement.
1
Continuous fog
Every time Continuous fog is applied every time.
Once
Once
Continuous fog is applied only once
before the 1st measurement.
Date/Time Set by ten-key display. Sets year, month, day, time (24hrs), minute and second.
Installation
date/time
Sph/Cyl step
0.12 Sph/Cyl is displayed by 0.12D step.
0.25
0.25 Sph/Cyl is displayed by 0.25D step.