62
SPECIFICATIONS AND PERFORMANCE
ELECTRIC RATING
Source voltage:100-240V AC, 50-60Hz
Power input: 75VA
SAFETY DESIGNATIONS PER IEC 60601-1 STANDARD
• Type of protection against electric shocks: Class I
The Class I equipment provides means to connect itself to the protective grounding system of
utilities to thereby independently provide protection against electric shocks by keeping con-
nectable metal components nonconductive in case of a failure in the basic insulation.
• Degree of protection against electric shocks: B type applied component
The B type applied component provides the specified degree of protection against electric
shocks with regard to the reliability particularly of leak current, patient measuring current and
protective utility connection (in case of Class I equipment).
• Degree of protection against harmful intrusion of water (IEC 60529): IPX0
This product does not provide protection against intrusion of water.
(The degree of protection against harmful ingress of water defined in IEC 60529 is IPX0)
• Classification by sterilization/disinfection method specified by manufacturer
This product does not have a component requiring sterilization/disinfection.
• Classification by safety of use in air/flammable anesthetic gas, oxygen or nitrous oxide/flamma-
ble anesthetic gas atmosphere
• Equipment not suited for use in air/flammable anesthetic gas, oxygen or nitrous oxide/flam-
mable anesthetic gas atmosphere
• This product should be used in an environment free of flammable anesthetic gas and other
flammable gases.
• Classification by operation mode
Continuous operation refers to an operation under normal load conditions, within the specified
temperature and without limitations on the operating time.
DIMENSIONS AND WEIGHT
Dimensions: 286~326mm(W) × 445~526mm(D) × 466~615mm(H)
Weight : 19.0kg