TARGET PRECISION
If better than normal precision is required, the 4640-B may be
programmed to automatically increase the count time to reach the
desired precision.
The 4640-B count time will be adjusted (up to a maximum of 60
min) to obtain the specified precision. The required precision time is
rounded to the nearest 15-second interval.
6. SPECIAL FUNCTIONS
Press SHIFT and SPECIAL. Either press YES once then
START/ENTER, or 5 for:
Target Precision
----.-- xxx
Input Precision
& Press ENTER
Input the desired precision and press START/ENTER.
The gauge will request the thickness of the overlay for
measurement. Input the thickness and press START/ENTER.
Place the gauge on the asphalt and start the 1-minute baseline count.
Target Precision
xx.xx
Time: xx mins.
(START to begin)
When the baseline count is complete, start the target precision
count.
NOTE
Refer to page B-2 for precision specifications.
Model 4640-B 6–9