ERASE STANDARD COUNTS
The Erase Standard Counts function allows the user or technician to
erase all four standard counts from gauge memory. This may be
desirable if the gauge has been repaired or mechanical changes have
been made. These factors may affect the standard counts and
comparing the new one to the average of the last four may cause the
gauge to fail the standard count test when nothing is wrong with the
gauge.
NOTE
This function will erase all the standard counts from
memory!
Press SHIFT and SPECIAL.
Either press YES four times then START/ENTER, or press
14 for the display:
Do you want to
erase all
Standard Counts
in gauge?
To erase the standard counts, press YES.