EasyManua.ls Logo

Vaisala CT25K - 5. Functional Description; 5.1 Theory of Operation

Vaisala CT25K
139 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Ceilometer CT25K
CT25K-U059en-2.1 User’s Guide
35
FEDCBA98 Alarm (A), Warning (W), and internal status information. Each character
is a hexadecimal representation of four bits, altogether 32 bits (b00-b31),
with the following breakdown. Interpretation as follows:
F: b31 Laser temperature shut-off (A)
b30 Laser failure (A)
b29 Receiver failure (A)
b28 Voltage failure (A)
E: b27 (spare) (A)
b26 (spare) (A)
b25 (spare) (A)
b24 (spare) (A)
D: b23 Window contaminated (W)
b22 Battery low (W)
b21 Laser power low (W)
b20 Laser temperature high or low (W)
C: b19 Internal temperature high or low (W)
b18 Voltage high or low (W)
b17 Relative Humidity is > 85 % (option) (W)
b16 Receiver optical cross-talk
compensation poor (W)
B: b15 Blower suspect (W)
b14 (spare) (W)
b13 (spare) (W)
b12 (spare) (W)
A: b11 Blower is ON
b10 Blower heater is ON
b09 Internal heater is ON
b08 Units are METERS if ON, else FEET
9: b07 Polling mode is ON
b06 Working from battery
b05 Single sequence mode is ON
b04 Manual settings are effective
8: b03 Tilt angle is > 45 degrees
b02 High background radiance
b01 Manual blower control
b00 (spare)
For example, if the battery voltage is too low, the internal heater is on and
units are meters, a warning is given and the second line appears as
0W ///// ///// ///// 00400300.
3RD LINE
L End-of-Text and CRLF