Ceilometer CT25K
User’s Guide CT25K-U059en-2.1
66
See Figure 5-11 and Table 5-2 for DMC50 DIP switch settings.
Figure 5-11 DIP Switch Settings of the DMC50
Table 5-2 DIP Switch Settings of the DMC50
Board Switch position Switch setting Function
DMC50 S2-1 OFF Disconnect digital GND
from CASE
S2-2 ON Enable watchdog timer
S2-3 OFF Disable download of operating
system
S2-4 ON Enable SRAM battery back-up