Ceilometer CT25K
CT25K-U059en-2.1 User’s Guide
47
CEILO>GET FACTORY
FACTORY
BEAMSPLITTER: 96%
**
IN LASER: 185 **
OUT LASER: 1200 **
COARSE COMP.: 13 **
FINE COMP.: 125 **
RECEIVER TEST VALUE: 350 **
CLEAN WINDOW: 200mV **
CEILO>
GET MESSAGE
MESSAGE
ANGLE CORRECTION: ON
HEIGHT OFFSET: 0
NOISE H2 COMPENSATION: OFF
PROFILE SCALE: 100%
PORT: DATA
TYPE: MSG1
UNITS: FEET
MODE: AUTOSEND
CEILO>GET OPER_MODE
OPERATION MODE: CONTINUOUS
CEILO>GET PORT
MAINTENANCE PORT BAUDS: 2400, E71
DATA PORT BAUDS: 300, E71
MODEM: CCITT (300)
MODEM STATUS: ON
YOU ARE USING: MAINTENANCE PORT
CEILO>GET UNIT_ID
UNIT ID: 0
NOTE
* Individual values are automatically adjusted by the
software.
**Individual original factory settings or updated original
settings in conjunction with subassembly replacement.
4.7 Manual angle setting
The automatic tilt angle measurement can be disabled and replaced with a
manual value. To set the manual angle value type :
CEILO>ADVANCED
PASSWORD ACCEPTED !
CEILO>SET MESSAGE ANGLE_MEAS MANUAL 5
MANUAL ANGLE: 5
To check the current state of tilt angle setting, get the status message :