Ceilometer CT25K
User’s Guide CT25K-U059en-2.1
42
4.4.11 Status Message "S"
The Status Message displays internal monitoring of the whole unit including
prevailing parameter values for voltages, receiver and transmitter,
temperatures, environmental factors, and internal heating. The message is
meant mainly for testing and maintenance purposes. The Status Message is
displayed by the command GET STATUS.
An example of the status message is presented below:
(
CTA38S0
J
↵
0W ///// ///// ///// 00400200↵
VOLTAGES (UNIT 0.1V)↵
P12 125 M12 -126 P5G 54 M5G -54 VCA 225↵
P13 128 M13 -124 P5R 50 M5R -50 BAT
*098↵
P18 178 PHV 2306 PFB 17 P65 674 CHA 144↵
↵
RECEIVER TRANSMITTER↵
GAIN H PLEN L↵
BAND N PQTY 64K↵
SAMP 10MHz OUT 1416mV↵
SENS OK SENS 101%↵
COMP 013 125 IN 190↵
↵
TEMPERATURES ENVIRONMENT↵
BLOWER +20C WINDOW 210mV 102%↵
CPU +34C RADIANCE +60mV↵
LASER +29C ANGLE +3DEG (M)↵
LENS +34C HUMIDITY NONE↵
OUTSIDE +23C↵
INHEATER ON OUTHEATER OFF BLOWER OFF↵
L
↵
NOTE
An asterisk (*) in front of a variable indicates that an
alarm or warning limit of that variable is exceeded. In the
example above the battery voltage is too low.
(M) is shown after the ANGLE value only if the manual
angle setting is used.
Interpretation of the status message is as follows:
1ST LINE
Identical to that of message no. 1 except that the second to last character
which identifies the message number, is always S.
2ND LINE
Identical to that of message no. 1