4.4.11 Status Message "S"
The Status Message displays internal monitoring of the whole unit including
prevailing parameter values for voltages, receiver and transmitter,
temperatures, environmental factors, and internal heating. The message is
meant mainly for testing and maintenance purposes. The Status Message is
displayed by the command GET STATUS.
An example of the status message is presented below:
(
CTA38S0
J
0W ///// ///// ///// 00400200
VOLTAGES (UNIT 0.1V)
P12 125 M12 -126 P5G 54 M5G -54 VCA 225
P13 128 M13 -124 P5R 50 M5R -50 BAT
*098
P18 178 PHV 2306 PFB 17 P65 674 CHA 144
RECEIVER TRANSMITTER
GAIN H PLEN L
BAND N PQTY 64K
SAMP 10MHz OUT 1416mV
SENS OK SENS 101%
COMP 013 125 IN 190
TEMPERATURES ENVIRONMENT
BLOWER +20C WINDOW 210mV 102%
CPU +34C RADIANCE +60mV
LASER +29C ANGLE +3DEG (M)
LENS +34C HUMIDITY NONE
OUTSIDE +23C
INHEATER ON OUTHEATER OFF BLOWER OFF
L
NOTE
An asterisk (*) in front of a variable indicates that an
alarm or warning limit of that variable is exceeded. In the
example above the battery voltage is too low.
(M) is shown after the ANGLE value only if the manual
angle setting is used.
Interpretation of the status message is as follows:
1ST LINE
Identical to that of message no. 1 except that the second to last character
which identifies the message number, is always S.
2ND LINE
Identical to that of message no. 1