The shortest output interval (with n = 0) outputs the measurement messages as quickly as the
transmitter’s internal measurement cycle produces them, without additional delay.
Example:
intv 1 min
Output interval: 1 MIN
7.7.4 Output a Reading Once
Use the SEND command to output a single measurement message. The output message is sent
in the format that has been set using the FORM command.
SEND [address]<cr>
Syntax Description
address
Address of the transmitter, range 0 ... 255.
Specifying the address is only necessary if the target transmitter
is in POLL mode, and has not been accessed with the OPEN
command.
Example:
send
Tdf= 16.71 'F Tdfatm= -6.19 'F H2O= 907 ppm
7.8
Configuring Measurement Parameters
7.8.1 Set Measurement Output Format
Use the serial line command FORM to change the measurement message sent by the
transmitter. You can freely define the output message to include the desired parameters,
formatting options, text strings, and additional fields.
FORM [modifier string]<cr>
Syntax
Description
modifier string
String of parameters and modifiers that defines the output
format, length 1 ... 153 characters. Maximum length may be
shorter when text strings are used.
Chapter 7 – Serial Communication
51