EasyManua.ls Logo

Vaisala DRYCAP DMT143 - Wiring and Power Supply

Vaisala DRYCAP DMT143
90 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Digital
Analog
Figure 6 DMT143 Digital and Analog Connectors
2
4
1 3
12 ... 28 VDC
RS-485 +
RS-485 -
GND
1
2
3
4
Figure 7 Wiring the Digital Connector
1
3
4
2
mA
1
3
4
2
V
18 ... 28 VDC
12 ... 28 VDC
2
4
1 3
Figure 8 Wiring the Analog Connector
4.4.1 Power Supply Requirements
The minimum supply voltage required by the DMT143 transmitter depends on the analog
output type that is used:
For voltage output: 12 ... 28 VDC.
For current output: 18 ... 28 VDC. The maximum load is 500 Ω.
The power supply should maintain the voltage for all load conditions.
If measuring in pressures above 20 bar
a
(290 psi
a
) or temperatures below 0 °C, the supply
voltage must be 24 ... 28 VDC.