If you have altered the DMT143 serial line settings and start using the MI70 hand-
held indicator, always switch o the other power supply (disconnect “Analog”
cable if in use) before powering up the MI70.
SERI [b p d s]<cr>
Syntax Description
b
baud rate (4800, 9600, or 19200)
p
parity (n = none, e = even, o = odd)
d
data bits (7 or 8)
s
stop bits (1 or 2)
Example (set serial line to 9600 baud, no parity, 7 data bits, and 1 stop bit):
seri 9600 e 7 1
SCI Baud P D S: 9600 E 7 1
More Information
‣
Operation with DM70 Handheld Dew Point Meter (page 38)
7.9.3 Set Transmitter Address
To operate in the POLL mode, the transmitter must have an address. If multiple transmitters
share the same serial line, each transmitter must have a dierent address.
ADDR [address]<cr>
Syntax
Description
address
Transmitter address, range 0 ... 255.
Example:
addr
Address : 0 ?
More Information
‣
Set Serial Line Operating Mode (page 57)
DMT143 User Guide M211435EN-J
58