7.10.3 Configure the Alarm LED
Use the ALARM command to configure the alarm LED functionality.
ALARM [LED] [Alarm] [Level]<cr>
Syntax Description
LED
Enable or disable the entire LED functionality.
Allowed values:
ON - Enable the LED functionality.
OFF - Turn o the LED completely.
Alarm
Enable or disable the dew point alarm.
Allowed values:
ON - Enable the dew point alarm.
OFF - Turn o the dew point alarm.
Level
Dew point temperature above which the alarm LED will turn red.
The LED will turn green when the measured dew point
temperature falls below this level.
Range -70 ... +60 °C T
df
.
Example (show current settings):
alarm ?
LED display : OFF
Alarm display : OFF
Setpoint Td(°C): -10.00
Example (change values line by line, enter to skip line):
alarm
LED display : OFF ? on
Alarm display : OFF ? on
Setpoint Td(°C): -10.00 ? -40
Example (change alarm level to -40 °C T
df
):
alarm on on -40
LED display : ON
Alarm display : ON
Setpoint Td(°C): -40.00
Example (turn o the dew point alarm only):
DMT143 User Guide M211435EN-J