RH oset and gain
T oset and gain
Tdf oset and gain
Example (shows default values, no user adjustment done):
l
RH offset : 0.00000000E+00
RH gain : 1.00000000E+00
T offset : 0.00000000E+00
T gain : 1.00000000E+00
Tdf offset: 0.00000000E+00
Tdf gain : 1.00000000E+00
Example (shows that a -1.2 °C Tdf oset has been applied by the user):
l
RH offset : 0.00000000E+00
RH gain : 1.00000000E+00
T offset : 0.00000000E+00
T gain : 1.00000000E+00
Tdf offset: -1.20000000E+00
Tdf gain : 1.00000000E+00
7.8.6 Set User Adjustment Parameters
Use the LI command to set the user adjustment parameters directly. This command is useful
for restoring the desired adjustments, and removing the eects of user adjustment without
having to perform the factory reset.
This command is provided primarily for applying an oset to the Tdf
measurement. Do not adjust the other parameters unless instructed to do so by
a Vaisala Service Center.
CAUTION!
LI<cr>
After entering the command, you will be prompted to enter new values one at a time. Press
ENTER without giving a value to move to the next parameter.
Example (shows a +0.8 °C Tdf oset being applied):
li
RH offset : 0.00000000E+00 ?
RH gain : 1.00000000E+00 ?
T offset : 0.00000000E+00 ?
T gain : 1.00000000E+00 ?
Tdf offset: 0.00000000E+00 ?
Tdf gain : 1.00000000E+00 ?
DMT143 User Guide M211435EN-J