Replace the filter as follows:
1. Turn the filter counterclockwise until it is loose.
2. Pull the filter straight out carefully.
Do not damage the sensor.CAUTION!
3. Take the new filter, and insert it to the filter thread.
4. Tighten the new filter to 5 Nm by turning it clockwise.
More Information
‣
Spare Parts and Accessories (page 73)
8.1.4 Field Check Using DM70 Handheld Dew Point Meter
Vaisala recommends the following equipment for field checking:
• A fully charged MI70 indicator (the measurement display that is included in the DM70
package).
• A calibrated reference probe that is suitable for the intended measurement environment
and dew point range. The DMP74B probe is recommended for checking the dew point
temperature measurement accuracy.
• You also need a connection cable (Vaisala order code 219980SP) to connect the DMT143
to the MI70 indicator.
Chapter 8 – Maintenance
63