Refer to Error Messages in Insight Software (page 67) for more information about the error states.
Table 16 Error Codes in MI70
Transmitter Error Code in MI70 Corresponding Error
1
hex
Temperature measurement error. [44]
2
hex
Humidity measurement error. [45]
40
hex
Device settings corrupted. [50]
80
hex
Additional configuration settings corrupted. [51]
100
hex
Sensor coecients corrupted. [52]
200
hex
Main configuration settings corrupted. [53]
800
hex
Supply voltage out of range. [55]
10000
hex
Non-volatile memory read/write failure. [57]
20000
hex
Firmware checksum mismatch. [49]
9.4 Unknown Serial Settings
You can check the current serial settings of your transmitter by connecting to the serial line, or
with Vaisala Insight software.
More Information
Accessing Serial Interface in Devices with Modbus Output (page 42)
Vaisala Insight Software (page 36)
DMT143 User Guide M211435EN-J