Register Description Data
Format
Unit / Valid
Range
Settings
795 031A
hex
Filtering factor 32-bit float 0.00001 … 1
Communication
1537 0600
hex
Address 16-bit
integer
1 ... 247
1538 0601
hex
Bit rate (b/s) 16-bit
integer
5 = 9600
6 = 19200
1539 0602
hex
Parity, data, stop bits 16-bit
integer
0 = N81
1 = N82
2 = E81
3 = E82
4 = O81
5 = O82
1540 0603
hex
Response delay 16-bit
integer
0 ... 1020 ms
1541 0604
hex
Interface mode 16-bit
integer
6 = MODBUS
1542 0605
hex
Restart device 16-bit
integer
1 = Restart the
device
LED Alarm
3073 0C00
hex
Indicator mode 16-bit
integer
0 = O
1 = Status only
2 = Alarm
3074 0C01
hex
Alarm parameter 16-bit
integer
6 = T
d/f
3075 0C02
hex
Alarm setpoint 32-bit float -70 °C ... +60 °C
Factory Settings
7681 1E00
hex
Restore settings 16-bit
integer
1 = Restore
factory settings
Appendix A – Modbus Reference