4.3 Insight Main View
Figure 13 Insight Main Menu and Settings
1 Select to access Insight main menu.
• Configure Device: environmental compensation settings, analog output 1 and 2
settings, Modbus and BACnet configuration, filtering factor, error limits and
general settings.
• Export Settings: creates a text file export of the device settings.
• Calibrate: options for calibrating and adjusting RH and T output, testing and
adjusting voltage (V) output levels, and restoring factory adjustments.
• Communication: contains a quick access selection for restarting the device.
• Factory Default Settings: restores the transmitter back to default settings, clears
any user adjustments and restores the latest factory calibration.
• About Device: general device information such as serial number and software
version.
2 Select Settings to switch between the Basic Mode and Advanced Mode user modes,
change the units of parameters (metric/non-metric), enter a factory code to access
restricted functionalities, or view information about the Insight software.
3 Monitoring provides options for monitoring and recording selected parameters, and
exporting the monitoring data as a CSV (comma-separated values) file.
4 Device information menu with the following tabs:
• Measurements: measurement graph view with parameter drop-down selection.
• Calibration Information: read-only information about the latest stored calibration.
• Diagnostics: troubleshooting and administrative information about the device
status. Also includes an option to export the device error log as a text file. When
contacting Vaisala support, it is recommended to include an up-to-date export of
the error log with the support request.
Chapter 4 – Operating with Insight PC Software