9. Technical Data
9.1 Specifications
Table 5 Relative Humidity Measurement Performance
Property Specification
Humidity sensor Vaisala HUMICAPâ R2
Measurement range 0 ... 100 %RH
Stability ±0.5 %RH/year in typical HVAC applications
Accuracy at 0 … +40 °C (+32 … +104 °F) (Incl. Non-linearity, Hysteresis, and Repeatability)
0 ... 90 %RH ±1.5 %RH
90 ... 100 %RH ±2.5 %RH
Accuracy at +40 … +80 °C (+104 … +176 °F) and -40 ... 0 °C (-40 ... +32 °F) (Incl. Non-linearity,
Hysteresis, and Repeatability)
0 … 90 %RH ±2.5 %RH
90 … 100 %RH ±3.5 %RH
Factory calibration uncertainty ±1.0 %RH
Start-up and Response Time
Start-up time at +20 °C (+68 °F) 8 s
Response time (T63) at +20 °C (+68 °F) 15 s
Calculated Humidity Parameters (Default Analog Output Scale)
Dew point -40 ... +80 °C (-40 ... +176 °F)
Dew point / frost point -40 ... +80 °C (-40 ... +176 °F)
Absolute humidity
0 … 300 g/m
3
(0 … 131.1 gr/ft
3
)
Wet bulb temperature -40 ... +80 °C (-40 ... +176 °F)
Enthalpy -40 … 1600 kJ/kg (-9.5 ... 695.6 Btu/lb)
Mixing ratio 0 … 600 g/kg (0 … 4200 gr/lb)
Table 6 Temperature Measurement Performance
Property Specification
Temperature sensor Pt1000 RTD Class F 0.1 IEC 60751
Measurement range -40 ... +80 °C (-40 ... +176 °F)