4.6 Changing Pressure Compensation Settings
with Insight
Figure 16 Pressure Compensation Settings
By default, the pressure value used when calculating HMD65 measurements is 1013.2 hPa.
To change the pressure compensation setting:
1. Select
> Configure Device.
2. To change the volatile compensation value (resets back to power-up default at device
restart), select Compensation setpoints, enter a value, and select Save.
3. To change the default compensation value that stays in use also after device restart, select
Compensation power-up defaults, enter a value, and select Save.
4. Select Close to exit the menu after you have saved the compensations.
More Information
Environmental Compensation (page 12)
HMD65 User Guide M212243EN-A