12. You will be prompted if you really want to adjust. Select Yes.
13. If the adjustment is successful, MI70 will show the text Adjustment Done, after which you
will return to the adjustment mode. At this point you can press Back and Exit to leave the
adjustment mode. The adjustment is now completed. If the adjustment cannot be applied,
MI70 will show the text Cannot adjust, possibly followed by a text stating the reason. A
possible reason for an adjustment failure is attempting to apply a very large correction to
the reading.