1. Introduction
6 Smart Plus (PHT-35LHS) Installation Manual
Installation Site
The PC Monitor, Emergency Cut Off Switch, and X-ray
Exposure Switch should be installed near the operator
so that he or she can manage them simultaneously in
an emergency.
Proper shielding of the room is essential. The installers
are responsible for verifying that all applicable radiation
safety requirements are met since the requirement
varies depending on the country. Since the requirement
varies depending on the country. It is the installer’s
responsibility to verify that all applicable radiation safety
requirements are met.
DO NOT install the equipment near other devices.
DO NOT install the equipment in an area that is
exposed to strong electromagnetic fields.
DO NOT install the equipment in an area where there is
the risk of an explosion.
The electrical installation of this system shall comply
with all local code requirements for electro-medical
systems: IEC 60364-7-710:2002.
It is strongly recommended that a UPS be installed at
the same time as the equipment.
The equipment, PC, and all peripheral devices must be
well-grounded.
Warnings Regarding X-ray Radiation
Failure to install this equipment in an approved location
may be dangerous to the patient and operator.
Stationary radiation shielding must be installed to
protect the operator from radiation.
The X-ray system may cause injury to the patient if
improperly used. Obey all federal and municipal
standards regarding radiation safety.
When exposing the patient to the X-ray, the operator
must be behind a protective wall or take other
protective actions. The operator should remain at least
2 m (7 feet) away from the X-ray when pressing the
Exposure Switch and observe the patient and capture-
progression.
Operators must provide protective clothing to the
patient before X-ray capturing. Pregnant women must
consult with a doctor before being exposed to an X-ray.