9. Installing Software
Smart Plus (PHT-35LHS) Installation Manual 133
Disabling the Packing Mode
Smart Plus has a unique feature─packing mode built into
the system to prevent the unit from being damaged while
shipping and transporting. Since it is in the packing mode
through the factory default, the packing mode is required to
be disabled at this step for a successful installation.
If the packing mode is not disabled, the equipment will not
be operated even after being turned on.
1. Click on the General tab on the Control Panel.
2. In the Networking option, enable the Serial checkbox and set Serial port and
Baud Rate as below, and then click Manager.
If an error has occurred during connection, make sure the
COM port setting is correct as follows:
1. Run the Device Manager.
2. Check the COM port assigned to “com0com” as shown in
the following figure.
3. If it is not assigned correctly, change the COM port in the
Networking option and click Save.
3. Enter the command PVER] to verify the current mode. Note that the equipment is
now in packing mode.