EasyManua.ls Logo

Viavi MAP-200 - Polarization Dependent Loss (PDL); Repeatability; IL Stability (maximum); Crosstalk (maximum)

Viavi MAP-200
103 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
mOSW-C1 Parameters
MAP-200 Optical Switch mOSW-C1 Installation and User Guide Standard
June 2016 Document Number 22112369-120 R000 Page 93
Parameter 1
1C Configuration
2D
Configuration
2E
Configuration
Polarization Dependent Loss (PDL)
6
Single Mode (SM) 0.04 dB 0.05 dB 0.07 dB
Repeatability
7
Sequential switching ±0.005 dB ±0.01 dB ±0.01 dB
Random switching ±0.025 dB ±0.04 dB ±0.04 dB
IL stability
8
(maximum)
Single Mode (SM) ±0.025 dB
Crosstalk (maximum)
Single Mode (SM) -80 dB
Multi Mode (MM) -60 dB
Optical Power
Max input power (optical) 300 mW
Lifetime
Lifetime 100 million
Switching time
≤ 24 ports
>24 ports
Electro-Mechanical (break to make) 20+10*(N-1) ms 55+30*(N-1) ms
Settling time to 90% final IL 60 ms 70 ms
Settling time for 99% final IL 90 ms 120 ms
Environment
Operation temperature 0 50 °C,
Operation humidity
15 to 80% RH,
0 to 40 °C non condensing
Storage temperature -30 to 60 °C
6
PDL tested at 1310nm and 1650nm.
7
Measured between two consecutive readings over 100 cycles.
8
Drift of any channel relative to reference channel at ±3°C deviation of ambient temperature over a 7-day period (168 hours).

Table of Contents