The table below shows the listing of the resistances requested for verification, the measurement current levels,
and the 95x specification at each current level. When prompted for each step the user should either –
• Press START to perform the verification.
• Press STOP to abort the entire sequence.
During each step, the user should note the displayed measurement on the 95x display and record the difference
between this measurement and the micro-ammeter reading, checking against the 95x specification shown in the
table below. When the user is satisfied that the step has been completed, the user should press START to continue
or STOP to abort the sequence.