95x Series Operating Manual - May 17, 2022
Page 91 of 155
The external voltage source is connected to one end of the material sample, with its common terminal grounded,
and the 95x RETURN terminal is connected to the other end of the material sample.
If the voltage being used is within the measurement capability of the 95x, then the output of the external voltage
source could also be connected to the HV terminal of the 95x, in which case the 95x display would also show the
actual applied voltage to the material sample.
A single step is used as follows
Seq 1 Step 1 DCI DC leakage required, so use DCI
TIME: user sec Use manual timing, press the START button when finished
Delay: 0.00sec Already connected, so no delay needed
Lim: 0n-5.000mA Set wide limit, the user will manually check the results
ARC DETECT:OFF Not required so disabled
ON FAIL: ABORT SEQ
SPECIFICATIONS
Specifications are valid at the 95x terminals for 1 year at ambient temperatures within ±5C of calibration
temperature (add 5% of accuracy specification per C outside of this). All specifications are relative to the
calibration standards used. Add ½ digit to all accuracies for displayed results (results available with enhanced
resolution from interfaces).
HV VOLTAGE MEASUREMENTS
DC Voltage Range 951i,952i,956i : 0.0 to +/-8000V
953i,954i,955i : 0.0 to +/-11000V
Option HSS or HSS-2 : 0.0 to +/-5000V
AC Voltage Range 951-4i,956i : 0.0 to 6000Vrms
955i : 0.0 to 10000Vrms
Option HSS or HSS-2 : 0.0 to 4000Vrms
DC Accuracy <(± 0.25% ± 0.5V)
AC Accuracy <(± 0.5% ± 0.5V ± 0.01% per Hz above 100Hz)
Input Impedance Nominally 100M
RETURN CURRENT MEASUREMENTS
Current Range 0.0nA to 200mA (DC or ACrms)
DC Current Accuracy <(± 0.25% ± 0.5nA)
AC Current Accuracy <(± 0.5% ± 20nA ± 0.005% per Hz above 100Hz)
Input Impedance <150uA : Nominally 5K
>150uA : Nominally 37
Arc Current 1 to 30mArms (50KHz-5MHz bandwidth)
<(± 10% ± 1mA) accuracy at 1MHz
4us, 10us, 15us, 20us, 30us or 40us measurement period
Cable Compensation <(± 1% of error)